22030710 TOC PREVENTATIVE MAINTENANCE AND
SERVICE TRAINING ON APOLLO 9000 TOC
Aim: Troubleshooting and maintenance of Apollo 9000 TOC
INTRODUCTION
TOC is determined as a measure of organic contamination for quality control and for
wastewater monitoring in...
More
.
22030710 TOC PREVENTATIVE MAINTENANCE AND
SERVICE TRAINING ON APOLLO 9000 TOC
Aim: Troubleshooting and maintenance of Apollo 9000 TOC
INTRODUCTION
TOC is determined as a measure of organic contamination for quality control and for
wastewater monitoring in the chemical, pharmaceutical, food and beverage industries.
Analysis of carbon is also essential for drinking water, groundwater, and wastewater
for compliance monitoring.
TROUBLESHOOTING APOLLO 9000 TOC
1.
High Baseline
No flow into the detector
Dirty combustion tube and catalyst
Dirty carrier gas supply
Incorrect zero adjustment on the NDIR
How do you lower the baseline?
No flow into the detector-make sure the gas flow is 200ml/min
Check at the gas inlet to the combustion and at the vent
Dirty combustion tube and catalyst
Dirty carrier gas supply
Connect the carrier- gas directly to the NDIR (make sure the gas flow is 200ml/min)
If the baseline drops the combustion tube and the catalyst need to be cleaned
If the baselin