Technical specifications
Order information
Order number 16288---TU
Power requirements 100–240 V AC
50/60 Hz
Maximum operating
pressure
9999 mbar|
145 psi
Minimum inlet
pressure
4000 mbar|
58 psi
Measuring ranges
Test pressure 100–8000 mbar|
1.
5–116 psi...
More
Technical specifications
Order information
Order number 16288---TU
Power requirements 100–240 V AC
50/60 Hz
Maximum operating
pressure
9999 mbar|
145 psi
Minimum inlet
pressure
4000 mbar|
58 psi
Measuring ranges
Test pressure 100–8000 mbar|
1.
5–116 psi
Pressure drop 1–2000 mbar|
0.
01–29 psi
System net volume
– with internal ref.
vessel 9000 ml
– with external ref.
vessel 100 l
Measuring accuracy
Pressure ± 0.
1% full scale
± 9.
5 mbar
Pressure drop ± 1 mbar
Volume determination ± 4%
Diffusion ± 5%
Water intrusion ± 5%
Bubble point ± 50 mbar|
0.
7 psi
Operating conditions
Ambient temperature +15 to + 35°C
Relative humidity 10–80%
Max distance between SC4
and multiunit (RS485) 100 m
Description
The Sartocheck 4 MultiUnit has been developed to enable parallel integrity testing of
multiple filters in the biopharmaceutical
industry.
The MultiUnit is an identical copy
of the Sartocheck 4, without the user interface and the data management system.
Each
MultiUnit connected to a Sartochec
Less